Global Medical Directory is part of the Informa Markets Division of Informa PLC

This site is operated by a business or businesses owned by Informa PLC and all copyright resides with them. Informa PLC's registered office is 5 Howick Place, London SW1P 1WG. Registered in England and Wales. Number 8860726.

syringe pump
Connect with this vendor - Learn about this product

Syringe pump

Model No. : SPA112 
Channel   Single 
Stackable   YES
Syringe Size: 5,10,20,30,50/60ml 
Infusion Modes: Rate Mode, Rate-Time, Rate-VTBI ,Time-VTBI, Body Weight 
Safety:  Double CPU to make sure higher standard infusion 
Drug Library: 20 drug list with drug code display 
Accuracy:  ±2% 
Time Preset: 00:01~99:59 (Hour:Min) 
Volume Range:  0~9999.9ml 
Flow Rate: Step by 0.1 ml/h when Rate <100ml/h, 1ml/h when Rate ≥100ml/h 
Flow Rate Range:
5ml Syringe   0.1ml/h-150ml/h 
10ml Syringe  0.1ml/h-300ml/h 
20ml Syringe  0.1ml/h-600ml/h 
30ml Syringe  0.1ml/h-900ml/h 
50/60ml Syringe  0..1ml/h-1200ml/h 
Purge/Bolus Rate/Top Flow Rate: 

5ml Syringe        150ml/h  5ml Syringe       150ml/h
10ml Syringe      300ml/h 10ml Syringe      400ml/h
20ml Syringe      600ml/h 20ml Syringe      600ml/h
30ml Syringe      900ml/h 30ml Syringe      900ml/h
50/60ml Syringe  1200ml/h 50/60ml Syringe  1800ml/h
Audible & Visible Alarms:
Auto Seft-Test, Syringe Dislocation, Occlusion, Near End,Syringe Empty, VTBI Completion,
Low Battery, Battery Exausted,  Motor Malfunction, Incorrect Syringe Specification,Circuit Malfunction, Master CPU Malfunction, Monitoring CPU Malfunction,Parameters Over Limitation, A.C Disconnection, A.C connection 
KVO: 0.1-5.0ml/h Adjustable 
Occlusion Pressure:

High: 40 KPa±20KPa 
Middle: 60 KPa±20KPa 
Low: 100KPa±20KPa 
Max Infusion Pressure: 120KPa 
Battery   ≥8 hours 
Power Consumption  30VA
Power Supply AC:100-240V,50Hz/60Hz 
Battery  Lithium battery,11.1/2000mAh 
Classification Class I,type CF,IPX4 
Dimension 26×21.5×11cm
Weight  2kg
Optional Function Ambulance DC: 12V