Global Medical Directory is part of the Informa Markets Division of Informa PLC

This site is operated by a business or businesses owned by Informa PLC and all copyright resides with them. Informa PLC's registered office is 5 Howick Place, London SW1P 1WG. Registered in England and Wales. Number 8860726.

DYMIND DH76 AUTOLOADER HEAMATOLOGY ANALYZER
Connect with this vendor - Learn about this product

DYMIND DH76

80tests/ h | DH-76

Tri-angle Laser scatter,chemical dye ,and flow cytometry;
5-part differentiation, 29 parameters;
1 3D scatter gram ,3 2D scatter grams and 3 histograms;
Only 20ul sampling volume;
Up to 80 samples per hour;
60 samples continously loading ;
3 counting modes: whole blood ,capillary blood and prediluted;
up to 100,000 samples results can be stored;
Bi-directional Lis connection.

Technical Specifications:

Principles
Impedance method for WBC, RBC, PLT counting
Cyanide-free reagent for Hemoglobin test
Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel

Parameters
WBC, Lym%, Lym#,Mon%, Mon#,Neu%, Neu#,Eos%, Eos#, Bas%, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT,and 4 research parameters P-LCC , P-LCR, LIC%, LIC#, ALY%, ALY#
1 3D and 3 2D 3 scatter grams and 3 histograms

Throughput
Up to 80 samples per hour

Sample Volume
Whole Blood: 20ul
Prediluted: 20ul

Test Mode
CBC CBC+DIFF

Performance
Carryover Repeatability Linearity
WBC ≤0.5% ≤2.0% (4-15×109/L) 0.00-300×109/L
RBC ≤0.5% ≤1.5% (3.5-6.0×1012/L) 0.00-8.5×1012/L
HGB ≤0.6% ≤1.5% (110-180g/L) 0-250g/L
PLT ≤1.0% ≤4.0% (150-500×109/L) 0.00-3000×109/L

Data Storage Capacity
Up to 100,000 results including numeric and graphical information

Communication
LAN port supports HL7 protocol

Operating Environment
Temperature: 15°C-30°C
Humidity: 30-85%
Air Pressure: 70-106 kPa

Power Requirement
A.C. 100-240V ≤300VA 50/60Hz

Dimension and Weight
620mm(L) × 620mm(W) × 535mm(H)

Specialty: